You Have Received Your Sample! Now What?

As part of our custom product manufacturing process, UTAK offers matrix testing to support Baseline Testing (MTX-BASELINE) prior to spiking analytes or beginning final packaging. This sample you received is a 10mL frozen sample of the starting matrix. Below you will find our recommended steps for testing the sample and completing the Baseline Testing (MTX-BASELINE) form.

Testing Procedure

  1. Allow the frozen matrix sample to thaw at room temperature with the cap on.
  2. Swirl gently for 3–4 minutes to ensure a homogeneous mixture.
  3. Swirl gently each time an aliquot is removed to maintain consistency.
  4. Test the matrix sample in the same manner as patient specimens, following the full instructions of your standard test method.
  5. Minimize the time the matrix sample remains at room temperature during handling to help preserve its integrity.

Approval Form Completion

  1. Indicate Accept or Reject based on whether the matrix meets your specifications.
  2. Record the recovered analyte values and indicate Accept or Reject accordingly.
  3. Sign and date the completed form and email it to approvals@utak.com.

If you have any questions or concerns please reach out to one of our Control Freaks for assistance!